Search In this Thesis
   Search In this Thesis  
العنوان
”The Realization of Luminous Flux Scale and Its Application in Preparing Calibrated and Tested Lamps”.
الناشر
Ain Shams University. Faculty of College for Women. Departmen of physics.
المؤلف
Haridy,Manal Ahmed Abdel Naby
هيئة الاعداد
مشرف / Mohmed Mamoun El-Ganainy
مشرف / da Badr El-Bialy
مشرف / Arnold Allan Gaertner
مشرف / Manal Ahmed Abdel Naby Haridy
تاريخ النشر
2007 .
عدد الصفحات
207P.
اللغة
الإنجليزية
الدرجة
الدكتوراه
التخصص
الفيزياء النووية والطاقة العالية
تاريخ الإجازة
1/1/2007
مكان الإجازة
جامعة عين شمس - كلية البنات - physics
الفهرس
Only 14 pages are availabe for public view

from 246

from 246

Abstract

In this work the realization of total luminous flux is being carried out through the calibration of incandescent transfer standard lamps directly from (NRC) primary standards of total luminous flux such lamps were used for an international comparison of the total luminous flux scales of five national institute.
The design, constriction and characterization of new temperature controlled photometer to be used with the NRC three meter integrating sphere in calibration of incandescent working standard lamps.
The final results of the uncertainties are
1-The relative expanded (k=2) uncertainty budget for the total luminous flux for the incandescent transfer standard lamps which is 2.15%.
2-The relative expanded (k=2) uncertainty budget for the total luminous flux for the incandescent working standard lamps which is 2.4%
3-The uncertainty budget for the spectral responsivity of the new temperature controlled photometer is mentioned in the text.